Devadason SG., Dore ND., LeSouëf PN., Wildhaber JH., Hayden MJ. Salbutamol delivery from a hydrofluoroalkane pressurized metered dose inhaler in pediatric ventilator circuits; an in-vitro study. Chest 1998;113:186-91 Abstract: STUDY OBJECTIVES: The aim of our study was to determine the in vitro delivery of salbutamol from a pressurized metered-dose inhaler (pMDI) containing hydrofluoroalkane (HFA) propellant through various delivery devices to four models of a pediatric lung. DESIGN: To determine the effect of electrostatic charge, delivery of salbutamol was initially assessed with a multistage liquid impinger (MSLI) through an inline nonchamber device (Baxter MDI Adapter) and a small (AeroChamber MV) and a large (Nebuhaler) inline chamber device. Following this, the delivery was assessed to four lung models appropriate for a child of 70 kg, 50 kg, 15 kg, and 4 kg, with the same three reduced static devices inserted directly into a pediatric ventilator circuit. MEASUREMENTS AND RESULTS: Reduction of electrostatic charge improved small particle delivery through holding chambers to the MSLI by 12 to 14%. In the ventilator model, the mean delivery was between 1.9% and 5.4% for the nonchamber device, between 14.3% and 27.2% for the small holding chamber, and between 7.2% and 25.7% for the large holding chamber. Delivery was the least efficient in the 4-kg model compared to the 70-kg, 50-kg, and 15-kg models. CONCLUSIONS: Salbutamol from an HFA pMDI is delivered efficiently through in-line holding chambers with reduced static in pediatric ventilator settings. A large holding chamber has no advantage over a small holding chamber. In addition, salbutamol delivery is more efficient through a holding chamber than through a nonchamber device. 1998 |